SCPI Commands Reference

266

IQS-12004B

Specific Commands

SOURce:ON

Description This command turns the source ON or OFF.

Syntax

SOUR:ON

Parameter

The parameter is a boolean value indicating the new

state of the source:
“1” - Sets the source to ON
“0” - Sets the source to OFF

Response

None

Example

SOUR:ON 1

Note

This is a blocking function i.e., In progress appears in the

status bar and you must wait before sending a new command.

SOURce:ON?

Description The response to this query indicates if the source was active.

Syntax

SOUR:ON?

Parameter

None

Response

A boolean value representing the current state of the source:
“1” - The source is ON
“0” - The source is OFF

Example

SOUR:ON?