COM Objects Reference

312

IQS-12004B

Test Parameters Interface—Properties

Test Parameters

AvailableTests

Description

The response to this query indicates the tests that can be

performed according to the modules detected, the calibration

and the reference measurement.

Syntax

object.AvailableTests

Parameters

None

Response

Enumeration EAvailableTests
For more information, see Public Enum EAvailableTests on

page 329.

Access

Get

Note

Depending on the module detected, you can perform an IL,

Il/ORL, IL/PDL, IL/ORL/PDL, and ORL measurement.

Test Parameters

KeepILForAllPolarization

Description

This command keeps the insertion loss curves for all states of

polarization (4), when a PDL acquisition has been performed.

Syntax

object.KeepForAllPolarization

Parameters

Boolean

Response

Boolean

Access

Get/Let

Note

Default: False
This command requires a lot of memory space as there are four