Using the Power Meter Linearity Verification Tool

DWDM Passive Component Test System

153

Connecting the Modules for Linearity Testing

The following diagram illustrates how to link the required modules.

Connect synchronization cables (the black ones) between:

³

The IQS-9401 Wavelength Reference Module (WRM) and the power
meter to be tested.

³

The power meter to be tested and the reference power meter.

You can use standard patchcords for optical connections. However, to
ensure more stability during the verification, you may prefer to use the
supplied semi-rigid patchcord to connect the optical source output port to
the In port of the Wavelength Reference Module.

IQS-1643T

Power Meter

IQS-9600

1X2 Coupler

IQS-1643T

Power Meter

IQS-2600CT

(1550nm)

IQS-3100

Variable

Attenuator

IQS-9401

WRM

2

1

C

IN

OUT

IN

OUT

Synchronization
cables

Calibrated power meter acting as a reference

Power meter(s)
to be tested

1x2 coupler, distributing optical power to the
DUT and the reference power meter