Automating or Remotely Controlling the System

192

IQS-12004B

GPIB Specific Commands—Quick Reference

SOP

<0 | 1 | 2 | 3 |>

This command indicates on which

SOP to take the IL curve, when a

PDL acquisition has been

performed.

SOP?

(0 | 1 | 2 | 3 |)

The response to this query indicates

the current SOP when reading the

IL during a PDL acquisition.

PSA

MAX

--

This command optimizes the

power.

PSA

MAX?

(0 | 1)

The response to this query indicates

if the power has been optimized.

SOP

<0 | 1 | 2 | 3 |>

This command sets a new

polarization state on the IQS-5150

polarization state adjuster.

SOP?

(0 | 1 | 2 | 3 |)

The response to this query indicates

the current polarization state of the

IQS-5150 polarization state adjuster.

SOUR

MINW?

(9999.999)

The response to this query indicates

the minimum wavelength (in nm)

of the current source.

MAXW?

(9999.999)

The response to this query indicates

the maximum wavelength (in nm)

of the current source.

ON

<0 | 1>

This command turns the source ON

or OFF.

ON?

(0 | 1)

The response to this query indicates

if the source was active.

STAB?

(Time)

The response to this query indicates

(in seconds) how long the source

has been turned on.

Command

Parameter/Response

Description