Testing Multiple DUTs with the LabVIEW Application

126

IQS-12004B

Calibrating the System Wavelength Response

To perform a wavelength response calibration:
1.
If the System menu is grayed out, that means no test has been created

since the application was started. Consequently, no calibration can be
performed until you start a new test. For more information, see Starting
a New Test
on page 118.

2. Select Wavelength Response Calibration from the System menu of

the main window.

3. Connect start fiber to the specified switch port and power meter

channel.

4. When you are satisfied with the value displayed in the Power Meter

Monitor box, click Start.

5. Repeat steps 3 and 4 for each power meter channel. The connection

information section provides you with the number of the switch port
and the number of the power meter channel indicating where the
patchcord must be connected.

6. When you are done with wavelength response calibration, simply click

Close to return to the main window.

Note: The zone labeled Instructions guides you through the whole wavelength

response calibration process.