Automating or Remotely Controlling the System

204

IQS-12004B

Configuring Your DWDM Passive Component Test System as a GPIB Controller

4. Restart the IQS-12004B controller unit. Once the unit has restarted, the

following dialog box appears.

5. Test the GPIB card as follows:

5a. Check the Do not show at Windows startup box located at the

bottom of the dialog box.

5b. Click Verify your hardware and software installation. The

NI-488.2 Troubleshooting wizard is automatically started.

5c. Tests take about 15 seconds to complete. Wait until the Status

field displays “passed”, and then close all windows.

The IQS-12004B controller unit can now be used as a GPIB controller.