beautypg.com

Calibration:orlmandrelrtj – EXFO IQS-12004B DWDM Passive Component Test System for IQS-500 User Manual

Page 260

background image

SCPI Commands Reference

248

IQS-12004B

Specific Commands

CALibration:ORLMandrelrtj

Description This command performs an ORL calibration with a mandrel

and reference test jumper.

Syntax

CAL:ORLM

Parameter

None

Response

None

Example

CAL:ORLM

CALibration:ORLMandrelrtj?

Description The response to this query indicates if an ORL calibration with a

mandrel and reference test jumper has been performed.

Syntax

CAL:ORLM?

Parameter

None

Response

A boolean response representing the state of the ORL mandrel

RTJ calibration acquisition:
“1” - The ORL mandrel RTJ calibration is performed
“0” - The ORL mandrel RTJ calibration is not performed

Example

CAL:ORLM?