Automating or Remotely Controlling the System

196

IQS-12004B

SCPI Error Messages

17

Unable to execute another command at this time.

Please wait until the current command is complete.

18

An inappropriate channel value has been assigned.

20

Unable to use the modules because the module

initialization has not been done. Please initialize the

modules.

21

Unable to perform an acquisition because the ORL

calibration with an RTJ has not been performed.

Please perform an ORL calibration.

22

Unable to perform an acquisition because the ORL

calibration with a mandreled RTJ has not been

performed. Please perform a calibration of the ORL

with a mandreled RTJ.

23

Unable to perform an acquisition because the

wavelength response calibration has not been

performed. Please perform a wavelength response

calibration.

24

Unable to perform an acquisition because the IL

reference has not been performed. Please performed

an IL reference.

25

Unable to perform an acquisition because the IL/PDL

reference has not been performed. Please performed

an IL/PDL reference.

26

Unable to perform an acquisition because the ORL

reference has not been performed. Please perform

an ORL reference.

27

Unable to get a curve because the acquisition result

curve is empty. Please perform an acquisition.

28

Unable to change the polarization state because the

power has not been optimized. Please optimize the

power.

Error Number

Description