Automating or Remotely Controlling the System
196
IQS-12004B
SCPI Error Messages
17
Unable to execute another command at this time.
Please wait until the current command is complete.
18
An inappropriate channel value has been assigned.
20
Unable to use the modules because the module
initialization has not been done. Please initialize the
modules.
21
Unable to perform an acquisition because the ORL
calibration with an RTJ has not been performed.
Please perform an ORL calibration.
22
Unable to perform an acquisition because the ORL
calibration with a mandreled RTJ has not been
performed. Please perform a calibration of the ORL
with a mandreled RTJ.
23
Unable to perform an acquisition because the
wavelength response calibration has not been
performed. Please perform a wavelength response
calibration.
24
Unable to perform an acquisition because the IL
reference has not been performed. Please performed
an IL reference.
25
Unable to perform an acquisition because the IL/PDL
reference has not been performed. Please performed
an IL/PDL reference.
26
Unable to perform an acquisition because the ORL
reference has not been performed. Please perform
an ORL reference.
27
Unable to get a curve because the acquisition result
curve is empty. Please perform an acquisition.
28
Unable to change the polarization state because the
power has not been optimized. Please optimize the
power.
Error Number
Description