beautypg.com

Verifying linearity on a specific channel, Mportant – EXFO IQS-12004B DWDM Passive Component Test System for IQS-500 User Manual

Page 172

background image

Using the Power Meter Linearity Verification Tool

160

IQS-12004B

Verifying Linearity on a Specific Channel

Verifying Linearity on a Specific Channel

Linearity must be verified on each channel of the power meter to be
tested. After each acquisition, a linearity value is calculated:

reference power meter result – DUT result = linearity value

The channel status (Pass/Fail) will be determined by comparing the
obtained value to the limit linearity (value appearing on the upper right of
the Instrument function tab).

The channel will be tagged as Pass if

otherwise, it will be identified as Fail.

I

MPORTANT

Linearity values obtained when power is under –55 dBm are not
taken into account when determining the channel status, due to
the power meter specifications.

Current
channel
status

Pass/Fail
criterion
(non-editable)

obtained linearity

permitted limit