EXFO IQS-12004B DWDM Passive Component Test System for IQS-500 User Manual
Page 220

Automating or Remotely Controlling the System
208
IQS-12004B
IQS-12004B DWDM Passive Component Test System (Standard) COM Error Messages
IQS-12004B DWDM Passive Component Test
System (Standard) COM Error Messages
Error Number
Description
ErrAcqBefore
Unable to get a curve because the acquisition result curve is
empty. Please perform an acquisition.
ErrAcqChannel
An inappropriate number of acquisition channels has been
assigned.
ErrAcqRun
Unable to start an acquisition at this time. An acquisition is
already in progress.
ErrAcqStopped
The acquisition has stopped.
ErrAcqType
Invalid acquisition type. The acquisition has stopped.
ErrCalOrlMandrelRtjBefore Unable to perform an acquisition because the ORL
calibration with a mandreled RTJ has not been performed.
Please perform a calibration of the ORL with a mandreled
RTJ.
ErrCalOrlRtjBefore
Unable to perform an acquisition because the ORL
calibration with an RTJ has not been performed. Please
perform an ORL calibration.
ErrCalWrBefore
Unable to perform an acquisition because the wavelength
response calibration has not been performed. Please
perform a wavelength response calibration.
ErrCalWrm
Invalid IQS-9401 calibration.
ErrChannelNumber
An inappropriate channel value has been assigned.
ErrCommandNot
SupportedBySource
The command is not supported by the current source
module.
ErrDetection
Unable to detect hardware module.
ErrHardNotCompatible
Unable to perform this test because some modules are not
selected or not installed.