Referencing the System with MultiPath Testing Option

114

IQS-12004B

Calibrating the System Wavelength Response

To perform a null measurement:
1.
Put protective caps on power meter detectors as well as on the input

and output ports of the WRM.

2. Perform the null measurement.

OR
1.
Connect the rigid and semi-rigid patchcords to the appropriate

modules (source, polarization state adjuster, wavelength reference
module and switch). See To work with the MultiPath Testing Option,
you will have to add a switch to your standard IQS-12004B DWDM
Passive Component Test System to benefit from its inherent features.
on
page 26 for more infor
mation.

2. Put protective caps on power meter detectors.

3. Turn the source off.

4. Perform the null measurement.

Calibrating the System Wavelength Response

This operation allows to correct any variations in the response and
sensitivity of the different power meters.

To perform a wavelength response calibration:
1.
Connect output port 1 of the switch to channel 1 of the power meter.

2. Perform the wavelength response calibration.

3. Repeat the previous steps with all of the remaining power meter

channels.