beautypg.com

Setiltrace, setorltrace, setpdltrace – EXFO IQS-12004B DWDM Passive Component Test System for IQS-500 User Manual

Page 368

background image

COM Objects Reference

356

IQS-12004B

Analysis Interface—Methods

Analysis

SetIlTrace, SetOrlTrace, SetPdlTrace

Description

Each of these methods allows to re-analyze a previously

acquired trace.

Syntax

object.SetIlTrace (Trace)
object.SetOrlTrace (Trace)
object.SetPdlTrace (Trace)

Parameters

Trace: Required. A Variant value containing a two-dimensional

array representing pairs of X,Y coordinates of the trace to

re-analyze.

Response

None

Note

The SetIlTrace, SetOrlTrace and SetPdlTrace methods let you

re-analyze any acquired trace at a later time. This is mainly

useful when you want to compare the results before and after

applying drift traces for environmental type tests.
Once the trace of interest is set, you must start the DoAnalysis

method before retrieving results.
Use this method in conjunction with the following methods:
AddRefExtZeroTrace, AddRefExtTrace, GetDriftTrace,
GetCorrectedIlTrace.