Thermcraft XST-6-0-36-3V User Manual
Page 247
Installation and Operation Handbook
Configuration
2416 Controller
6-15
Name
Description
Values
Meaning
CAL
Calibration
In this mode you can
1. Calibrate the instrument using a mV source -
rcAL
or ref source
cal.
2. Offset the calibration to account for errors in actual sensor
measurement and a ref sensor -
UCAL
or user calibration
3. Return to factory set calibration -
FACT
or factory set calibration.
rcAL
Calibration
point
nonE
No calibration
PV
Calibrate main Process Value input.
PV.2
Calibrate DC input, or PV 2.(not
2416)
1A.Hi
Calibrate DC output high - Module 1
1A.Lo
Calibrate DC output low - Module 1
2A.Hi
Calibrate DC output high - Module 2
2A.Lo
Calibrate DC output low - Module 2
3A.Hi
Calibrate DC output high - Module 3
3A.Lo
Calibrate DC output low - Module 3
I
NPUT
C
ALIBRATION
For ‘
CAL
’ = ‘
PV’, or ‘PV.2
’, the following parameters apply.
PV
PV Calibration Value
IdLE
Idle
mv.L
Select 0mV as the calibration point
mv.H
Select 50mV as the calibration point
V 0
Select 0Volt as the calibration point
1. Select calibration value
V 10
Select 10V as the calibration point
2. Apply specified input
CJC
Select 0
o
C CJC calibration point
3. Press to step to ‘
GO’ rtd
Select 400
Ω as the calibration point
HI 0
High impedance: 0Volt cal’n point
HI 1.0
High impedance: 1.0 Volt cal’n point
FACt
Restore factory calibration
GO
Start calibration
no
Waiting to calibrate PV point
Select ‘
YES’ with
or
YES
Start calibration
Wait for calibration to
buSy
Busy calibrating
complete.
donE
PV input calibration completed
FAIL
Calibration failed
Go to User
calibration
table - See
also chapter 7
Go to input
Calibration table
Go to
DC Output
Calibration
table