Programmer Operation

Installation and Operation Handbook

5-12

2416 Controller

Segment type

Select the segment type using

or

:

rmP.r

Ramp to a new setpoint at a set rate

rmP.t

Ramp to a new setpoint in a set time

dwEl

Dwell for a set time

StEP

Step to a new setpoint

cALL

Call another program as a subroutine

(only available in 4-program controllers)

End

Make this segment end of program.

Press

The parameters that follow

tYPE’ depend on the type of segment selected.

Parameter

Segment type selected

rmP.r rmP.t dwEl StEP cALL

End

Hb

9

9

9

9

tGt

9

9

9

rAtE

9

dur

9

9

PrG.n

9

outn

9

9

9

9

9

cYc.n

9

dwEl

9

End.t

9

Pwr

9

Table 5-3 Parameters that follow segment

tYPE

Holdback type
[Only appears when Holdback per segment has been selected.]

Use

or

to select:

OFF

: Holdback disabled

Lo

:

Deviation Low Holdback

Hi

:

Deviation High Holdback

bAnd

:

Deviation Band Holdback

Press

Target setpoint

Target setpoint for ‘

rmP.r’, ‘rmP.t’ or ‘StEP’ segments.

Set the target setpoint using

or

.


Press

Continued on the next page.

Hb

bAnd

tgt

100

type

rmp.r