Operation

Installation and Operation Handbook

2-16

2416 Controller

2.7 ALARMS

2.7.1 Alarm

annunciation

Alarms are flashed as messages in the Home display. A new alarm is displayed as a double
flash followed by a pause, old (acknowledged) alarms as a single flash followed by a pause.
If there is more than one alarm condition, the display cycles through all the relevant alarm
messages. Table 2-1 and Table 2-2 list all of the possible alarm messages and their meanings.

2.7.2

Alarm acknowledgement and resetting

Pressing both

and

at the same time will acknowledge any new alarms and reset any

latched alarms.

2.7.3 Alarm

modes

Alarms will have been set up to operate in one of several modes, either:
Non-latching, which means that the alarm will reset automatically when the Process

Value is no longer in the alarm condition.

Latching, which means that the alarm message will continue to flash even if the alarm

condition no longer exists and will only clear when reset.

Blocking, which means that the alarm will only become active after it has first entered a

safe state on power-up.

2.7.4 Alarm

types

There are two types of alarm: Process alarms and Diagnostic alarms.

2.7.5 Process

alarms

These warn that there is a problem with the process which the controller is trying to control.













* In place of the dash, the first character will indicate the alarm number.

Table 2-1 Process alarms

Alarm

Display

What it means

_

FSL*

PV Full Scale Low alarm

_

FSH*

PV Full Scale High alarm

_

dEv*

PV Deviation Band alarm

_

dHi*

PV Deviation High alarm

_

dLo*

PV Deviation Low alarm

_

LCr*

Load Current Low alarm

_

HCr*

Load Current High alarm

Alarm

Display

What it means

_

FL2*

Not available in 2416

_

FH2*

Not available in 2416

_

LOP*

Working Output Low alarm

_

HOP*

Working Output High alarm

_

LSP*

Working Setpoint Low alarm

_

HSP*

Working Setpoint High alarm

4rAt

PV Rate of change alarm
Always assigned to Alarm 4