User Calibration
2408 and 2404 Controller
7-3
OFFSET CALIBRATION
Offset calibration is used to apply a single fixed offset over the full display range of the
controller.
To calibrate, proceed as follows:
1. Connect the input of the controller to the source device to which you wish to calibrate.
2. Set the source to the desired calibration value.
3. The controller will display the current measurement of the value.
4. If the displayed value is correct, then the controller is correctly calibrated and no further
action is necessary. If it is incorrect, then follow the steps shown below.
Select ‘
FuLL’ access level, as described in Chapter 3.
x 3
Input list header
Press
until you reach the input list header.
Press
until you reach the ‘
CAL’ display.
Calibration type
•
FACt:
Factory
Calibration
•
USEr:
User
Calibration
Use
or
to select ‘
FACt’.
Selecting ‘
FACt’ reinstates the factory calibration and allows the
application of a single fixed offset.
Press
continued
on the next page
Displayed Value
Input
Factory Calibration
Fixed Offset
CAL
FACt