Thermcraft XST-6-0-36-3V User Manual
Page 134
User Calibration
Installation and Operation Handbook
7-8
2408 and 2404 Controller
CALIBRATION POINTS AND CALIBRATION OFFSETS
If you wish to see the points at which the User calibration was performed and the value of the
offsets introduced, then these are shown in Configuration, in ‘
CAL-Conf’.
The parameters are:
Name
Parameter description
Meaning
pt1.L
Low calibration point for Input 1
The factory calibration point at which the low
point offset was performed.
pt1.H
High calibration point for Input 1
The factory calibration point at which the high
point offset was performed.
OF1.L
Offset Low for Input 1
Calculated offset, in display units.
OF1.H
Offset High for Input 1
Calculated offset, in display units.
pt2.L
Low calibration point for Input 2
The factory calibration point at which the low
point offset was performed.
pt2.H
High calibration point for Input 2
The factory calibration point at which the high
point offset was performed.
OF2.L
Offset Low for Input 2
Calculated offset, in display units.
OF2.H
Offset High for Input 2
Calculated offset, in display units.
Note: The value of each of the parameters in the above table may also be altered by using
the
/
buttons.