Metrohm 799 GPT Titrino User Manual
Page 96
2.13. Method memory, keys
799 GPT Titrino
92
>backup
dir.name:
Backup of the internal method memory on the
card
(directory name with up to 10 characters)
Existing methods in this directory will be deleted, and
all methods from the internal method memory are
stored in the directory.
>reload
dir.name:
Reload methods from the card to the internal
method memory
(directory name with up to 10 characters)
Primary all methods from the internal method memory
are deleted. Afterwards all methods from the directory
of the card are copied into the internal method
memory.
>format
card label
format: no
Format the card
Card label (with up to 8 characters)
The card label will be printed in the card report and in
the full result report (if the card is inserted, see page
81).
Confirmation of formatting the card (yes, no)
Upon formatting the card, all data on the card are
deleted.
After formatting a new card, the date for battery change
must be entered.
>change battery
date
Battery change date
Date for changing the battery (YYYY-MM-DD)
see page 93.
Possibilities of the card
Methods can be stored on the card.
• Simple exchange of methods between different users, laboratories, factories. Every
user can use his card with his own methods. The user specific card label will be
printed in the full result report, if the card remains inserted.
• Methods can be stored in different directories, e.g. in directories depending on the
sample type or the user.
• The card can be used as an extended memory.
• You will find many application methods on the application card 6.6047.000.