2 parameters for set – Metrohm 799 GPT Titrino User Manual
Page 29
![background image](/manuals/586575/29/background.png)
2.6 Parameters, key
799 GPT Titrino
25
2.6.2 Parameters for SET
SET1, SET2:
Control parameters for EP1 and EP2.
titration parameters
are valid for the global titration sequence.
stop conditions:
Parameters for the termination of the titration.
statistics
:
Calculation of mean values and standard deviation,
see page 77.
preselections
:
ON/OFF of various auxiliary functions such as auto-
matic requests after the start and activate pulse.
>SET1
EP at pH OFF
**titr.
dynamics OFF
**titr.
max.rate 10.0 ml/min
**titr.
Control parameters for EP1 or EP2, resp.
Preset EP1 (input range depends on the measured
quantity:
pH: 0...±20.00,
OFF
U, Ipol: 0...±2000 mV, OFF
Upol:
0...±200.0 µA, OFF)
If EP1 is "OFF", no further inquiries under SET1 appear.
Distance from EP where constant dosing should stop
and controlling begins. (control range, input range
depends on the measured quantity:
pH: 0.01...20.00,
OFF
U, Ipol: 1...2000 mV, OFF
Upol:
0.1...200.0 µA, OFF)
"OFF" means largest control range, i.e. low titration.
Outside the control range, dispensing is performed
continuously, see also page 31.
Maximum dosing rate (0.01...150 mL/min, max.)
This parameter determines primarily the addition rate
outside the control range, see also page 31.
The maximum rate depends on the Exchange Unit:
Exchange Unit
max.
5 mL
15 mL/min
10 mL
30 mL/min
20 mL
60 mL/min
50 mL
150 mL/min
parameters
>SET1
>SET2
>titration parameters
>stop conditions
>statistics
>preselections