Metrohm 799 GPT Titrino User Manual
Page 161
3.2 Remote control commands
799 GPT Titrino
157
3.2.2.72. Mode.Parameter.Calibration.SmplChanger ON, OFF
Parameters for CAL: Calibration at Titrino.
With "ON", there are no hold points in the calibration sequence for entries, the
first buffer is measured directly.
3.2.2.73. Mode.Parameter.Calibration.ActPulse first,
all, OFF
Parameters for CAL: Output of a pulse on the I/O line "Activate", see page
224.
3.2.2.74. Mode.Parameter.Sequence.X.Select
method, pause, L4 output,
L6 output, info, prep, stirrer, OFF
Parameters for TIP: Selection of an element for step X (X = 1...30). For the
parameters of the elements see 3.2.2.75.
3.2.2.75. Mode.Parameter.Sequence.X.Method Method
name
Mode.Parameter.Sequence.X.Pause
0...999 999, INF
Mode.Parameter.Sequence.X.L4Output
active, inactive,
pulse, OFF
Mode.Parameter.Sequence.X.L6Output
as for L4
Mode.Parameter.Sequence.X.Info
up to 16 ASCII characters
Mode.Parameter.Sequence.X.Prep
internal D0, external D1,
external
D2
Mode.Parameter.Sequence.X.Stirrer
ON, OFF
Parameters for TIP: Parameters of the elements of TIP.
.Method: Method name of a method available in the user memory or on the
card. Up to 8 ASCII characters.
.Pause:
Pause time in s. INF means infinite. Continue the sequence with
&m $G.
.L4 Output: Warning: A pulse triggered by the limit value monitoring at L4 (pin
3) in a submethod sets an output set to active in TIP to inactive.
.L6 Output: Warning: An activate pulse at L6 output (pin 1) in a submethod
sets an output set to active in TIP to inactive.
.Info:
Entry of a message which is written into the display. The se-
quence remains in the display with the corresponding message.
Continue with &m $G.
.Prep:
Preparation of titrating buret.
.Stirrer:
Switching stirrer on/off.
3.2.2.76. Mode.Parameter.Presel.MeasMode
pH, U, Ipol, Upol, OFF
Mode.Parameter.Presel.MeasInput
1, 2, diff.
Mode.Parameter.Presel.Ipol 0...1...±127
Mode.Parameter.Presel.Upol 0...400...±1270
Mode.Parameter.Presel.PolElectrTest ON, OFF
Mode.Parameter.Presel.Temp -170....25.0...500.0
Parameters for TIP: Selection of the measured quantity for manual measure-
ments in the inactive state, see 3.2.2.2. Selection of the measuring input
(MeasInput) applies to measured quantities pH and U. "diff." means differen-
tial amplifier, see page 215. With Ipol the requests for the polarization current