7external inputs and outputs, 1 test of the remote interface – Metrohm 799 GPT Titrino User Manual

Page 203

background image

4.2.

Diagnosis

799 GPT Titrino

199

• Press .

diagnose press key 0...9

• Remove cable and resistor switch-box.

7

External inputs and outputs

This test is meaningful only if the 799 GDP Titrino is used interconnected with other instruments via the 'Remote' con-

nection. In addition, a 3.496.8550 test plug normally used in the repair service is required for this test. However, this

plug can also be purchased by customers under the above number.

For the sake of completeness, the procedure is described here. If a diagnostic test of the external inputs and outputs

is not required, continue with point 8.

5 18 4 17 3 16 1

2

6

7

8 13 19 20

12 24 11 23 10 22 9 21

Fig. 4

Connections in the 3.496.8550 test plug

• Press <4>

extern input/output test

7.1 Test of the remote interface

• Press <1>.

extern input/output test

• Press .

I/O-test-connector?

• Insert the 3.496.8550 test plug in port B 'Remote'. (Do not switch off instrument!)

• Press .

Test

sequence:

1. In case of an error an error message is displayed. If for example no test plug is con-

nected, error message error 50 01HEX

appears).

2. If no error is found, after about 1 s display shows

extern input/output o.k.