beautypg.com

Metrohm 799 GPT Titrino User Manual

Page 21

background image

2.6 Parameters, key

799 GPT Titrino

17

pause 0 s

**titr.




dos.element: internal D0




meas.input: 1






I(pol) 1 µA

U(pol) 400 mV


electrode test: OFF





temperature 25.0 °C

Waiting time (0...999 999 s)

Waiting time, e.g. for equilibration of the electrode after

the start or reaction time after dosing of start volume.

The pause can be aborted with .

Selection of the dosing element (internal D0, external

D1, external D2)

internal D0: internal buret of the Titrino

external D1/2: buret D1, D2, resp.

Measuring input for pH and U (1, 2, diff.)

Request for measuring input for pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 215.

With polarized electrodes, instead of the measuring

input the

polarization current (-127...127 µA)

or the

polarization voltage (-1270...1270 mV, in steps of 10

mV)

is inquired.

Electrode test (OFF, ON)

Test for polarized electrodes. Performed on

changeover from the inactive standby state to a

measurement. "OFF" means that the test is not

performed.

Titration temperature (-170.0...500.0 °C)

The temperature is continuously measured if a T

sensor is connected. This parameter is used for

temperature compensation in pH titrations.


>stop conditions



stop V: abs.

**titr.






stop V 99.99 ml

**titr.


factor 999999

**titr.







Stop conditions for the titration

If several stop conditions have been set, the criterion

which is met first applies.

Type of stop volume (abs., rel., OFF)

"abs.": absolute stop volume in mL.

"rel.": relative stop volume to sample size.

"OFF": stop volume switched off. Stop volume is not

monitored.

If "abs." is set:

Absolute stop volume (0...9999.99 mL)

If "rel." is set:

Factor for relative stop volume (0...

±

999 999)

Calculated as: Stop V (in mL) = factor ∗sample size