beautypg.com

Metrohm 799 GPT Titrino User Manual

Page 178

background image

3.2. Remote control commands

799 GPT Titrino

174

L6

Activate pulse + can be set in TIP

L7

Buret volume pulses

L8, 9

---

L10-13 Monitoring in modes STAT, DOS, DOC

.ResetLines: Lines are set to the inactive status (= high).

3.2.2.158. Assembly.Stirrer ON, OFF

Switching stirrer ON/OFF.

3.2.2.159. Setup.Comport

1, 2, 1&2

Selects the Titrino COM for the output of automatic info:

&Setup.Keycode

&Setup.Trace

&Setup.SendMeas

&Setup.AutoInfo

3.2.2.160. Setup.Keycode ON, OFF

ON means the key code of a key pressed on the Titrino is outputted. The key

code comprises 2 ASCII characters; table of the keys with their code, see

page 195. A keystroke of key 11 is sent as follows:

#11

The beginning of the message is marked by a space (ASCII 32).

3.2.2.161. Setup.Tree.Short ON, OFF

Setup.Tree.ChangedOnly ON, OFF

Definition of the type of answer to $Q.

.Short:

With "ON", each path is sent with only the necessary amount of

characters in order to be unequivocal (printed in bold in this manual). A com-

bination of .Short and .ChangedOnly is not possible.

.ChangedOnly:

Sends only the changed values, i.e. values which have

been edited. All paths are sent absolute, i.e. from the root.

3.2.2.162. Setup.Trace ON, OFF

The Titrino automatically reports when a value has been confirmed with

ter> at the Titrino. Message, e.g.:

&SmplData.OFFSilo.Id1"Trace"

The beginning of the message is marked by a space (ASCII 32).

3.2.2.163. Setup.Lock.Keyboard ON, OFF

Setup.Lock.Config ON, OFF

Setup.Lock.Parameter ON, OFF

Setup.Lock.SmplData ON, OFF

Setup.Lock.UserMeth.Recall ON, OFF

Setup.Lock.UserMeth.Store ON, OFF

Setup.Lock.UserMeth.Delete ON, OFF