3.1 General rules
799 GPT Titrino
115
E203
No Oven parameters: Oven not (correctly) connected.
Exit: The error message disappears on next start. If you don't wish oven
parameters in your report, select &Mode.Parameter.Presel.Oven "no" in your
method(s).
E205
Calibration interval is expired.
Exit: The error message disappears on next calibration or if you delete the
calibration.
E212
Transmission error from Remote Box. Unknown characters.
Exit: Rectify error and switch Titrino off and on again.
E213
Time-out error from PC keyboard (Remote Box)
Exit: Rectify error and switch Titrino off and on again.
E214
Check Remote Box. Remote Box not (properly) connected but activated in
&Config.Periph.RemoteBox.
Exit: Rectify error and switch Titrino off and on again.
E270
Overload in dosing element: piston is blocked.
Exit:
dos driver and press
E282
Overload in dosing element: cock is blocked.
Exit:
dos driver and press