2.6 Parameters, key

799 GPT Titrino

53

Measuring point list and monitoring

General information on measuring point list:
• Measured points are entered in the measuring point list at the preset time interval.
• If one (or more) entries appear during the refilling time, one measured point is entered

immediately after the refilling. The time reference otherwise remains unchanged,

however.

• If the stop time occurs within the refilling time period, after filling one measured point is

entered and only then the determination is stopped.

• The measuring point list always includes the values of time and volume. If monitoring

of measured values and/or temperature is active, the corresponding values are

entered too.

• The "mess." column shows a message if a limit value was exceeded in the last time

interval, see example below. "* " is used to mark if dosing was interrupted, e.g. owing

to refilling, action "wait" or "hold" on limit value violation or manual .

• The volume of the last measured point entry can differ from the final volume as time

may have elapsed between the last entry and the actual stop. The final volume is

available as variable C41 and can be used to calculate results.

Example of a measuring point list. Monitoring of measured value and temperature active.

More than 1 monitored quantity violated.

Temperature limit violated.

pH limit violated.

Dosing interruption.

'mp
799 GPT Titrino 01102 799.0010
date 2001-09-14 time 08:54 14
pH(init) 7.42 STAT pH ********
smpl size 1.0021 g
t/s V/mL pH T/°C mess.
0 0.0000 7.434 35.5 limit
30 0.9140 3.039 35.7 limit
60 1.0010 3.056 36.0 temp.
90 1.0780 3.022 36.2
120 1.1470 2.972 36.4 meas
150 1.2250 3.018 36.5 *
:
: