Metrohm 799 GPT Titrino User Manual
Page 168
3.2. Remote control commands
799 GPT Titrino
164
3.2.2.119. Config.DosPrep.DX.Dosino.DosTubing.Length 1...40...999.9
Config.DosPrep.DX.Dosino.DosTubing.Diameter 0.1...2...9.9
Config.DosPrep.DX.Dosino.AspirTubing.Length 1...25...999.9
Config.DosPrep.DX.Dosino.AspirTubing.Diameter 0.1...2...9.9
.DosTubing: Dosing tube. Length in cm, diameter in mm.
.AspirTubing: Aspiration tube. Length in cm, diameter in mm.
Start of sequence see 3.2.2.152.
3.2.2.120. Config.DosPrep.DX.Dosino.DosRate 0.01...150,
max.
Config.DosPrep.DX.Dosino.FillRate 0.01...150,
max.
Dosing and filling rate in ml/min for preparation. Start of sequence see
3.2.2.152.
3.2.2.121. SmplData.Status ON, OFF
On/off switching of silo memory. When the silo memory is switched on, the
sample data are fetched from the lowest valid silo line.
3.2.2.122. SmplData.OFFSilo.Id1
up to 8 ASCII characters
SmplData.OFFSilo.Id2
up to 8 ASCII characters
SmplData.OFFSilo.Id3
up to 8 ASCII characters
SmplData.OFFSilo.ValSmpl
6-digits, sign and decimal point
SmplData.OFFSilo.UnitSmpl
up to 5 ASCII characters
Current sample data.
The identifications Id1...Id3 can be used in formulas as sample-specific calcu-
lation constants C21...C23.
If "no unit" is desired for the unit of the sample size, the blank string must be
entered.
3.2.2.123. SmplData.ONSilo.Counter.MaxLines read
only
SmplData.ONSilo.Counter.FirstLine read
only
SmplData.ONSilo.Counter.LastLine read
only
Information on silo memory.
.MaxLines: Maximum possible number of silo lines.
.FirstLine: Lowest valid silo line.
.LastLine: Last occupied silo line.
3.2.2.124. SmplData.ONSilo.EditLine.1.Method
up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.Id1
up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.Id2
up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.Id3
up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.ValSmpl 6-digits, sign and dec.point
SmplData.ONSilo.EditLine.1.UnitSmpl
up to 5 ASCII characters
SmplData.ONSilo.EditLine.1.C24 read
only
SmplData.ONSilo.EditLine.1.C25 read
only
SmplData.ONSilo.EditLine.1.Mark read
only
etc., up to .255
Contents of a silo line.