3.2 Remote control commands
799 GPT Titrino
167
3.2.2.132. Info.PrepData.D0.Date read
only
Info.PrepData.D0.Time read
only
Info.PrepData.D1.Type read
only
Info.PrepData.D1.Date read
only
Info.PrepData.D1.Time read
only
identical to .D2
Data from the last correctly executed preparation.
3.2.2.133. Info.Checksums $G
Info.Checksums.MPList read
only
Info.Checksums.ActualMethod read
only
The checksums can be used to identify the content of a file unequivocally,
e.g. files with identical content
have identical results of the checksums. An empty file has checksum "0". The
calculation of the checksums is triggered with $G.
.MPList: Result of the checksum of the current measuring point list.
.ActualMethod: Result of the checksum of the current method in the working
memory. Identical methods with different method names have the
same results of the checksum.
3.2.2.134. Info.DetermData $G
Info.DetermData.Write ON,
OFF
Info.DetermData.MPList.1.Attribute read
only/read+write
Info.DetermData.ExV read
only/read+write
Info.DetermData.MPList.1.X read
only/read+write
Info.DetermData.MPList.1.Y read
only/read+write
Info.DetermData.MPList.1.Z1 read
only/read+write
Info.DetermData.MPList.1.Z2 read
only/read+write
for every measuring point
Determination data in hexadecimal format. A measuring point list is available
in mode DET, MET, STAT, DOS, DOC, SET, KFT, and MEAS.
Recalculation of the measuring data is triggered with $G.
.Write:
With "ON", the following nodes can be overwritten:
&Info.DetermData.MP.List, &Info.TitrResults.Var.C4X (X = 0...5),
&Info.TitrResults.TempVar.C7X (X = 0...9), and &Mode.Name.
.ExV:
Volume of the exchange unit, with which the determination was
executed
.MPList.1.Attribute: Attribute
.MPList.X: X coordinate, time
.MPList.Y: Y coordinate, volume
.MPList.Z1: Z1 coordinate, measuring value
.MPList.Z2: Z2 coordinate, temperature
3.2.2.135. Info.TitrResults.RS.1.Value read
only
etc., up to .9
Info.TitrResults.EP.1.V read
only
Info.TitrResults.EP.1.Meas read
only
etc., up to .2
Info.TitrResults.Var.C40 read
only/read+write
etc., up to .C47