3.2. Remote control commands

799 GPT Titrino

162

3.2.2.102. Config.Aux.RunNo

0...9999

Current sample number.

Set to 0 on power on and initialization. After 9999, counting starts again at 0.

3.2.2.103. Config.Aux.AutoStart 1...9999, OFF

Number of automatic, internal starts.

3.2.2.104. Config.Aux.StartDelay

0...999 999

Start delay time in s. During this time, the data of the preceding determination

are retained.

3.2.2.105. Config.Aux.ResDisplay bold, standard

Character set for the result display at the end of the determination.

3.2.2.106. Config.Aux.DevName

up to 8 ASCII characters

Name of the instrument for connections with several units. It is advisable to

use only the letters A...Z (ASCII No. 65...90), a...z (ASCII No. 97...122) and

the numbers 0...9 (ASCII No. 48...57) when the function Setup.AutoInfo

(3.2.2.169) is used at the same time.

If a name has been entered, it will be printed out in the result report (full,

short).

3.2.2.107. Config.Aux.Prog read

only

Output of the program version.

The Titrino sends "799.0010" on requests with $Q.

3.2.2.108. Config.RSSet1 $G

Config.RSSet1.Baud

300, 600, 1200, 2400, 4800, 9600,

19200, 38400, 57600, 115200

Config.RSSet1.DataBit 7,

8

Config.RSSet1.StopBit

1, 2

Config.RSSet1.Parity even,

odd, none

Config.RSSet1.Handsh HWs, SWchar, SWline, none

$G sets all RS settings. The changes are performed only if the instrument is

inactive. After the setting of the interface parameters, wait at least 2 s to allow

the components to equilibrate.

Settings of the values for the data transmission via the RS interface: baud

rate, data bit, stop bit, parity and type of handshake, see also page 180 ff.

Baud rates >9600 need a PC which is equipped accordingly (e.g. with 16550

component).

The setting of the values must be initiated with $G immediately after entry of

the values.