![background image](/manuals/586575/130/background.png)
3.2. Remote control commands
799 GPT Titrino
126
*Parameter
Tree part "Parameters for DOC"
├
.
DosPara
Dosing parameters
│ ├
.BeginMeas
Begin of measuring
depends on meas.quant. 3.2.2.57.
│ ├
.EndMeas
End of measuring
depends on meas.quant.
ditto
│ ├
.UnitMeas
Unit of measured quantity
read only
ditto
│ ├
.SweepTime Sweep
time
0...300...999 999
3.2.2.58.
│ ├
.Dyn
Dynamics
depends on meas.quant. 3.2.2.59.
│ ├
.UnitDyn
Unit of dynamics
read only
ditto
│ ├
.MaxRate
Maximum dosing rate
0.01... 150, max. ditto
│ ├
.MinRate
Minimum dosing rate
0.01...5.0...9999 ditto
│ ├
.Direction
Direction +,
-,
auto 3.2.2.60.
│ ├
.StartV Start
volume
│ │ ├
.Type
Type of start volume
abs., rel., OFF 3.2.2.61.
│ │ ├
.V
Volume for absolute start volume
0...999.99 ditto
│ │ ├
.Factor
Factor for relative start volume
0...±999 999
ditto
│ │ ├
.Rate
Dispensing rate for start volume
0.01...150.0, max. ditto
│ ├
.Pause
Waiting time after start volume
0...999 999
ditto
│ ├
.TDelta
Time interv. for meas.acquisition
1...10...999 999
3.2.2.52.
│ ├
.DosUnit
Selection of the dosing unit
internal D0, external D1,
│ │
external
D2
3.2.2.17.
│ ├
.MeasInput Measuring
input
1, 2, diff.
3.2.2.62.
│ ├
.Ipol
Polarization
current
0...1...±127 ditto
│ ├
.Upol
Polarization voltage
0...400...±1270 ditto
│ ├
.PolElectrTest
Test for polarized electrodes
ON, OFF ditto
│ ├
.Temp
Titration temperature
-170.0...25.0...500.0 3.2.2.53.
│ │
├
.
StopCond
Stop conditions
│ ├
.VStop
Stop volume
│ │ ├
.Type
Type of stop volume
abs., rel., OFF
3.2.2.20.
│ │ ├
.V
Volume for absolute stop volume
0...99.99...9999.99 ditto
│ │ ├
.Factor
Factor for relative stop volume
0...±999 999 ditto
│ ├
.FillRate Filling
rate
0.01...150.0,
max. 3.2.2.23.
│ │
├
.
Statistics
Statistics
│ ├
.Status
Status of statistics calculation
ON, OFF 3.2.2.24.
│ ├
.MeanN
No. of individual determinations
2...20 ditto
│ ├
.ResTab Result
table
│ │ ├
.Select
original,delete n,delete all
ditto
│ │ ├
.DelN
Deletion of individual results
1...20 ditto
│ │
├
.
Monitoring
Monitoring
│ ├
.MeasVal
Monitoring of measured values
│ │ ├
.Status
Status
ON, OFF 3.2.2.45.
│ │ ├
.LowLim Lower
limit
depends on meas.quant.
ditto
│ │ ├
.UpLim Upper
limit
depends on meas.quant.
ditto
│ │ ├
.Action
Action if out of limits
end, hold, wait, none ditto
│ ├
.Temp
Monitoring of temperatures
│ │ ├
.Status
Status
ON, OFF 3.2.2.47.
│ │ ├
.LowLim Lower
limit -170.0...500.0 ditto
│ │ ├
.UpLim Upper
limit
-170.0...500.0 ditto
│ │ ├
.Action
Action if out of limits
end, hold, wait, none ditto