2.6. Parameters, key

799 GPT Titrino

50

Sequence for STAT

After the start, the activate pulse is output and the stirrer

switched on.

The start delay time is waited off.

The sample identifications and the sample size are re-

quested.

The start volume is dispensed (without controlling or

measured value acquisition and there is no monitoring

during this time) and the pause is waited off.

Then the other start conditions are worked off in the order

given below:

Start time

Start measured value

Start rate

During this time the controlling is active as well as the

monitoring. However, there is no acquisition of measured

values.

Once the start conditions have been met, measured val-

ues (time, volume) are acquired and stored in the measur-

ing point list. If monitoring of measured values and/or

temperature is active, these values will also be stored in

the measuring point list. If limits are violated it will be

marked in the measuring point list.

If a T sensor is connected, the temperature is measured

continuously (in intervals of 2 s). In pH titrations, the pH

will be corrected as a function of temperature.

Without T sensor the values are measured without inter-

rupt by T measurements. The manually set temperature is

then valid.

The titration is terminated according to the first stop crite-

rion which has been met.

Stirrer is switched off.

Evaluations and calculations are carried out.

The measuring point list is the base for evaluations (and

curves).

Data are output.

(Activate pulse)

(Stirrer ON)

(Start delay)

(Start conditions)

(Request ident.)

(Request smpl size)

Controlling

Acquisition of meas.

(Monitoring)

Stop conditions

Calculations

(Stirrer OFF)

Data output