beautypg.com

Metrohm 799 GPT Titrino User Manual

Page 180

background image

3.2. Remote control commands

799 GPT Titrino

176

.V: Volume

.Meas:

Measured value associated to the cycle number.

The unit "assembly" must be preset (see 3.2.2.166).

3.2.2.168. Setup.SendMeas.Titrator.CyclNo ON, OFF

Setup.SendMeas.Titrator.V ON, OFF

Setup.SendMeas.Titrator.Meas ON, OFF

Setup.SendMeas.Titrator.dVdt ON, OFF

Setup.SendMeas.Titrator.dMeasdt ON, OFF

Setup.SendMeas.Titrator.dMeasdV ON, OFF

Setup.SendMeas.Titrator.ERC ON, OFF

Setup.SendMeas.Titrator.T ON, OFF

Selection of the values from the titrator which are sent in the set time interval

(see 3.2.2.165, formats see 3.2.2.144):

.CyclNo: Cycle number. Together with the cycle time (3.2.2.149), a time

frame can be set up. The other data belong to the corresponding

cycle number. The cycle number is set to 0 at the start of a

method and it is incremented until the end of the method.

.V: Volume.

.Meas:

Measuring value with STAT, DOS, DOC with activated measured

value monitoring

.dVdt:

associated volume drift.

.dMeasdt: associated measured value drift.

.dMeasdV: associated 1st derivative of the titration curve.

.ERC:

ERC in DET.

.T:

In STAT, DOS, DOC, with activated temperature monitoring.

The unit "titrator" must be preset (see 3.2.2.166).

3.2.2.169. Setup.AutoInfo.Status ON, OFF

Setup.AutoInfo.P ON, OFF

Setup.AutoInfo.T.R ON, OFF

Setup.AutoInfo.T.G ON, OFF

Setup.AutoInfo.T.GC ON, OFF

Setup.AutoInfo.T.S ON, OFF

Setup.AutoInfo.T.B ON, OFF

Setup.AutoInfo.T.F ON, OFF

Setup.AutoInfo.T.E ON, OFF

Setup.AutoInfo.T.H ON, OFF

Setup.AutoInfo.T.C ON, OFF

Setup.AutoInfo.T.O ON, OFF

Setup.AutoInfo.T.N ON, OFF

Setup.AutoInfo.T.Re ON, OFF

Setup.AutoInfo.T.Si ON, OFF

Setup.AutoInfo.T.M ON, OFF

Setup.AutoInfo.T.EP ON, OFF

Setup.AutoInfo.T.RC ON, OFF

Setup.AutoInfo.C.B1 ON, OFF

Setup.AutoInfo.C.R1 ON, OFF

Setup.AutoInfo.C.B2 ON, OFF

Setup.AutoInfo.C.R2 ON, OFF

Setup.AutoInfo.I ON, OFF