beautypg.com

Metrohm 799 GPT Titrino User Manual

Page 56

background image

2.6. Parameters, key

799 GPT Titrino

52

Stop times

Stop time: The stop time runs over the whole controlling time of the determination: It

starts after dosing of the start volume and after the pause:

The stop time can be absolute (in s) or relative to the sample size (time (in s)

= factor ∗smpl size).

t(delta):

t(delta) starts after controlling point (EP) has been reached the first time.

t(delay):

t(delay) can be used as "after reaction time". It starts after the last dosing

step, i.e. with every dosing, t(delay) will be restarted from zero.

V/mL

t/s

Last

dosing

Control point (EP)

reached the first time

t(delay)

t(delta)

stop time