2.6 Parameters, key

799 GPT Titrino

45

start rate OFF ml/min





time interval 2 s




titr.direction: auto







dos.element: internal D0




meas.input: 1






I(pol) 1 µA

U(pol) 400 mV


electrode test: OFF





temperature 25.0 °C


Start rate for data acquisition (0.01...150 mL/min, OFF)

sets "OFF".

Measured values will be acquired only if the current

rate is below the start rate. This criterion effective 10 s

after the start.

Time interval for data acquisition (1..999 999 s)

Time interval for the entry of the values in the

measuring point list. The measuring point list can

contain max. 500 points.

Direction is set automatically (+, -, auto)

auto: The direction is set automatically by the Titrino

(sign [U

init

- EP]).

+:

Direction of higher pH, higher voltage (more

"positive"), larger currents.

- :

Direction of lower pH, lower voltage, smaller

currents.

Selection of dosing element (internal D0, external D1,

external D2)

internal D0: internal buret of the Titrino

external D1/2: buret D1, D2, resp.

Measuring input for pH and U. (1, 2, diff.)

Request only with measured quantities pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 216.

With polarized electrodes, instead of the measuring

input the

polarization current (-127...127 µA)

or the

polarization potential (-1270...1270 mV, in steps of 10

mV)

is inquired.

Electrode test (OFF, ON)

Test for polarized electrodes. Performed on

changeover from the inactive standby state to a

measurement. "OFF" means that the test is not

performed.

Temperature (-170.0...500.0 °C)

If a T sensor is attached, the temperature is measured

continuously and the pH values corrected accordingly.

The last measured value is entered as the

"temperature" parameter.