Metrohm 799 GPT Titrino User Manual
Page 76
2.6. Parameters, key
799 GPT Titrino
72
activate pulse: OFF
hold points in the calibration sequence for inputs.
Calibration temperature and pH values of the buffers
(which are temperature dependent) must therefore be
entered in advance. The inputs in key are
valid.
Pulse output on the line "activate" (L6, pin 1) of the
remote socket (all, first, OFF)
See page 224.
Calibration sequence
After the start, the activate pulse is output.
The start delay time is waited off.
Then, the calibration temperature is measured.
It no T sensor is connected, you enter the temperature
manually. Store the value with
Enter the nominal value of the first buffer.
Store the value with
The first buffer is measured.
Enter the nominal value of the second buffer.
Store the value with
Leave the calibration with
The second buffer is measured.
As many buffers appear as have been specified in the
key (up to 9).
You may leave the calibration any time with
Data are output.
The calibration data are available for calculation:
C46: pHas
C47: Electrode slope
Calibration data can be viewed at any time with the
using the key sequence
(Start delay)
Measuring cal.temp.
or
entry
Buffer 1 pH
Buffer 1 pH
etc.
Measuring buffer 1
Measuring buffer 2
Data output
(Activate pulse)