Metrohm 799 GPT Titrino User Manual
Page 62
2.6. Parameters, key
799 GPT Titrino
58
Same possibilities for lines L11, L12, L13.
Pin assignment on socket "Remote":
L10 Pin
8
L11 Pin
13
L12 Pin
19
L13 Pin
20
>preselections
req.ident: OFF
req.smpl size: OFF
limit smpl size: OFF
low lim. 0.0
up lim. 999999
activate pulse: OFF
Preselections for the sequence
Request of sample identifications after start of titration
(id1, id1&2, all, OFF)
After start, sample identifications can be inquired
automatically: Only id1, id1 & id2, all three id's or no
inquiries.
Request of sample size after start of titration (value, unit,
all, OFF)
Limiting value check for sample size (ON, OFF)
With "on" the error message "sample size out." appears
if the entry is outside the set limits. The limiting values
are shown in the display window.
The absolute value of the limit is checked during sam-
ple size input and during the calculation of the results.
If "on" has been set:
Lower limit for sample size (0.0...999 999)
Upper limit for sample size (0.0...999 999)
Pulse output on line "activate" (L6, pin 1) of the remote
socket (ON, OFF)
see page 224.