2.6 Parameters, key

799 GPT Titrino

28


>stop conditions



stop V: abs.

**titr.




stop V 99.99 ml

**titr.


factor 999999

**titr.



filling rate max. ml/min

Stop conditions for titration

If this is not "normal", i.e. after reaching the EP.

Type of stop volume (abs., rel., OFF)

"abs.":

absolute stop volume in mL.

"rel.":

relative stop volume to sample size.

"OFF":

stop volume switched off. Stop volume

is not monitored.

If "abs." is set:

Absolute stop volume (0...9999.99 mL)

If "rel." is set:

Factor for relative stop volume (0...±999 999)

Calculated as: Stop V (in mL) = factor∗sample size

Filling rate after the titration

(0.01...150 mL/min, max.)

sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min


>preselections

conditioning: OFF





display drift: ON

cond.


drift corr: OFF

cond.



drift value 0.0 µl/min

cond.


req.ident: OFF

cond.








Preselections for the sequence

Automatic conditioning of titration vessel. (ON, OFF)

If conditioning is "on", between the titrations the titration

solution is constantly maintained at the (1st) end point.

When conditioning is performed, the volume drift can

be displayed during the conditioning:

Display of drift during conditioning (ON, OFF).

Volume drift.

Type of drift correction (auto, man. OFF)

auto: The value of the drift will be stored automatically

at the start of the titration.

Value for manual drift correction (0...99.9 µL/min)

Request of identifications after start of titration (id1,

id1&2, all, OFF)

After start, sample identifications can be requested

automatically: Only id1, id1 & id2, all three id's or no

inquiries.