175

Setup.Lock.Display ON, OFF

ON means disable the corresponding function:

.Keyboard: Disable all keys of the Titrinos

.Config:

Disable the key

.Parameter: Disable the key

.SmplData: Disable the key

.UserMeth.Recall: Disable "recall" in key

.UserMeth.Store: Disable "store" in key

.UserMeth.Delete: Disable "delete" in key

.Display: Disable the display, i.e. it will not be written to by the device pro-

gram of the Titrino and can be operated from the computer.

3.2.2.164. Setup.Mode.StartWait ON, OFF

Setup.Mode.FinWait ON, OFF

Holding points in the method sequence. If they are "ON", the sequence stops

until "OFF" is sent. Switching the instrument on sets both nodes to OFF:

.StartWait: Holding point right after starting a method or submethod in TIP

(holding point after AutoInfo !".T.GC").

.FinWait: Holding point at the end a method or submethod in TIP (holding

point after AutoInfo !".T.F").

3.2.2.165. Setup.SendMeas.SendStatus ON, OFF

Setup.SendMeas.Interval 0.08...4...16200, MPList

.SendStatus: ON means the automatic transmission of measured values (see

3.2.2.167 and 3.2.2.168) in the inputted interval is active.

.Interval: Time interval (in s) for the automatic transmission of associated

measured values defined under points 3.2.2.167 and 3.2.2.168.

The inputted value is rounded off to a multiple of 0.08. The small-

est possible time interval depends on the number of measured

values which have to be sent, on the baud rate, on the load on the

interface and on the type of device connection. With "MPList" the

measured values are sent at the time of their entry into the meas-

ured point list.

The automatic transmission is switched on/off with 'SendStatus'.

3.2.2.166. Setup.SendMeas.Select Assembly,

Titrator

Selection of the unit of which the measured values should be sent (3.2.2.167

or 3.2.2.168).

3.2.2.167. Setup.SendMeas.Assembly.CyclNo ON, OFF

Setup.SendMeas.Assembly.V ON, OFF

Setup.SendMeas.Assembly.Meas ON, OFF

Selection of the values from Assembly for the output in the set time interval

(see 3.2.2.165):

.CyclNo: Cycle number of the potential measurement. Together with the

cycle time (3.2.2.149), a time frame can be set up.

The cycle number is set to 0 on switching on the instrument and it

is always incremented as long as the instrument remains