beautypg.com

Metrohm 799 GPT Titrino User Manual

Page 7

background image

1.2 Controls and parts

799 GPT Titrino

3

Rear view of instrument:

7

RS232

interfaces

2 separate interfaces

for the connection of

printer, balance, and

computer


8

Remote

lines

(input/output)

for the connection of

the Remote Box,

Sample Changers,

robots etc.

9

Connection of electrodes and temperature sensor
• 2 high-impedance measuring inputs for pH and U measurements. They can ei-

ther be used separately or for differential potentiometry, see page 216.

Important: If you work with both measuring inputs in the same vessel, the same

reference electrode must be used.

• 1 measuring input for polarized electrodes, e.g. KF electrode
• 1 measuring input for PT100 or Pt1000 temperature sensor

10

Connection for separate keypad

11

Connection for power cable

With power supplies where the voltage is subject to severe HF disturbances, the Ti-

trino should be operated via an additional power filter, e.g. Metrohm 615 model.


12

Connection for external dosing units D1 and D2

685 Dosimats or 700 Dosinos


13

Connection for stirrer

728 Magnetic Stirrer, 802 Rod Stirrer, 703 or 727 Ti Stand
Supply voltage: 10 VDC (I ≤ 200 mA)


14

Rating

plate

with fabrication, series and instrument number