3.2 Remote control commands
799 GPT Titrino
169
etc. up to .9
The current values of the statistics calculation.
$Q sends, e.g.
ActN: Current value of the individual results
"3"
Data for MN1:
Mean: Mean value (decimal places as in result)
"3.421"
Std: Standard deviation (1 decimal place more than in result)
"0.0231"
RelStd: Relative standard deviation (in %, 2 decimal places)
"0.14"
3.2.2.139. Info.SiloCalc.C24.Name read
only
Info.SiloCalc.C24.Value read
only
Info.SiloCalc.C24.Unit read
only
for
.C25 as for .C24
Info.SiloCalc.C26.ActN read
only
Info.SiloCalc.C26.Mean read
only
Info.SiloCalc.C26.Std read
only
Info.SiloCalc.C26.RelStd read
only
for
.C27 as for .C26
The current values from the silo calculations. C26 is the mean value out of the
C24 variables; C27 comes from C25.
$Q sends:
C24.Name: Name of the assigned value
"RS1"
C24.Value: Value
"2.222"
C24.Unit: Unit of the assigned value
"%"
C26.ActN: Number of single results
"3"
C26.Mean: Mean (decimal places as for the result itself)
"3.421"
C26.Std: Standard deviation (decimal places as for the result + 1) "0.0231"
C26.RelStd: Relative standard deviation (in %, 2 decimal places)
"0.14"
3.2.2.140. Info.ActualInfo.Inputs.Status read
only
Info.ActualInfo.Inputs.Change read
only
Info.ActualInfo.Inputs.Clear $G
Info.ActualInfo.Outputs.Status read
only
Info.ActualInfo.Outputs.Change read
only
Info.ActualInfo.Outputs.Clear $G
Status sends the current status of the I/O lines, Change sends the information
regarding whether a change in status of a line has taken place since the last
clearing, Clear clears the change information. For the output, there is a con-
version from binary to decimal, e.g.
| 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 1 | 0 |
Line No.
| 13| 12| 11| 10| 9 | 8 | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |
Output: 2
1
+ 2
3
= "10"
1 means ON or change; 0 means OFF or no change.
The lines are assigned as follows (see also pages 222ff):
Inputs: Outputs:
0
Start (pin 21)
0
Ready (pin 5)
1
Stop (pin 9)
1
Cond. ok (pin 18)
2
Enter (pin 22)
2
Titration (pin 4)
3
Clear (pin 10)
3
EOD (pin 17)
4
Smpl Ready (pin 23)
4
Monitoring, line L4 (pin 3)
5
pin 11
5
Error (pin 16)
6
pin 24
6
Activate, line L6 (pin 1)