4.2. Diagnosis

799 GPT Titrino

196

4

Cylinder code, date, time

• Insert exchange unit or dummy to the internal dosing unit D0 and put the burette tip

into a collecting receptacle.

• If available, connect the external dosing units D1 or D2 with the exchange units.

• Press <0>.

date/time

cylinder code

• Press .

date XX-XX-XX

XX:XX:XX

code:D0 XX ml

activated dosing unit

mL-code

• Check date and time.

• Check whether the ml-code does correspond with the exchange unit placed or not.

For the sake of completeness, different exchange units can be inserted to check their ml code.

If desired, the ex-

change unit can be removed again. If no exchange unit is inserted display does not show the ml-code but „check

exchange unit !“.

• If external dosing unit are connected, select the dosing unit in question with key

<š>.

Check the external dosing unit in the same way as before the internal. If no dosing unit is connected display does not

show the ml-code but „check drive unit !“.

• Press .

diagnose press key 0...9

5

Motor timer test

• Press <6>.

motor-timer test

• Press .

pot.meter dV/dt  10?

• Turn knob ‘dV/dt’ to the right stop and press .

Test

sequence:

a) In a first step, the frequency of the RC oscillator (analogue rate) is tested over a period

of 1 second.

b) In a second step, the frequency of the quartz oscillator (digital rate) is tested over a

period of 1 second.

c) If no error is found, after about 5 s it appears

motor-timer test o.k.