Metrohm 799 GPT Titrino User Manual
Page 175

3.2 Remote control commands
799 GPT Titrino
171
ERC
34
- - - - - - -
T
(°C)
25.9 25.9 -
-
25.9 25.9 25.9 -
NV: Not Valid. If in the signal drift is OFF in modes MEAS and CAL, the signal
drift is NV.
OV will be sent for "overrange".
A time frame can be set up from the cycle number and the cycle time (see
3.2.2.149). The cycle number is set to 0 at the start of a method and it is in-
cremented until the end of the method.
3.2.2.145. Info.ActualInfo.MeasPt.Index read
only
Info.ActualInfo.MeasPt.X read
only
Info.ActualInfo.MeasPt.Y read
only
Info.ActualInfo.MeasPt.Z1 read
only
Info.ActualInfo.MeasPt.Z2 read
only
Info.ActualInfo.EP.Index read
only
Info.ActualInfo.EP.X read
only
Info.ActualInfo.EP.Y read
only
$Q sends the last entry into the measuring point list (.MeasPt) or the last entry
into the list of EP's with DET, MET, SET, KFT.
.MeasPt.X"165" Volume (DET, MET), time (STAT, DOS, DOC, SET, KFT,
MEAS), resp. of the MPList in s
.MeasPt.Y"3.654" Measured value (DET, MET), volume (STAT, DOS, DOC,
SET, KFT), resp. of the MPList in mL
.MeasPt.Z1"6.34" Measured value (STAT, DOS, DOC, SET, KFT, MEAS) of the
MPList, format depends on the measured quantity
.MeasPt.Z2"25.8" Temperature of the MPList in °C
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP
3.2.2.146. Info.ActualInfo.Oven.HeatTime read
only
Info.ActualInfo.Oven.SampleTemp read
only
Info.ActualInfo.Oven.LowTemp read
only
Info.ActualInfo.Oven.HighTemp read
only
Info.ActualInfo.Oven.GasFlow read
only
Info.ActualInfo.Oven.UnitFlow read
only
$Q sends the current values from a connected KF Oven. If no Oven is con-
nected, the values are empty.
.HeatTime: Heating time of sample in s.
.SampleTemp: Nominal sample temperature in °C.
.LowTemp: Lowest temperature during the sample heating time in °C.
.HighTemp: Highest temperature during the sample heating time in °C.
.GasFlow: Average gas flow during sample heating time.
.UnitFlow: Unit of gas flow.
3.2.2.147. Info.ActualInfo.Display.L1
up to 32 ASCII characters
Info.ActualInfo.Display.L8
up to 32 ASCII characters
Info.ActualInfo.Display.DelAll $G
Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
1. Lock the display, see 3.2.2.163.
2. Delete the whole display (.DelAll).