3.2. Remote control commands
799 GPT Titrino
148
Mode.Parameter.Evaluation.Recognition.Window.1.LowLim
pH:
0...±20.00, OFF
U,
Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
Mode.Parameter.Evaluation.Recognition.Window.1.UpLim
Input range as LowLim
etc. up to 9 windows
Mode.Parameter.Evaluation.FixEP.1.Value pH:
0...±20.00, OFF
U;
Ipol:
0...±2000, OFF
etc. up to 9 fix EP's
Upol: 0...±200.0, OFF
Mode.Parameter.Evaluation.pK ON,
OFF
Parameters for DET and MET: Evaluation of the EP's, see page 21.
.EPC: EP criterion in pH, in mV (with U and Ipol) or in uA (with Upol).
.Recognition.Selected: EP recognition.
all: All endpoints found are recognized.
great: Only the largest EP is recognized.
last: Only the last EP is recognized.
window: Only EP's that lie within set windows are recognized.
OFF: The EP evaluation is switched off.
.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U
and Ipol) or uA (with Upol).
.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and
Ipol) or uA (with Upol).
Windows are opened until the lower limit is set to OFF. For every
expected EP, an individual window must be set, see page 23.
.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. uA (for Upol). Fix EP's are
evaluated until the setting OFF is found.
.pK: pK or HNP evaluation. Possible only in pH and U titrations.
3.2.2.26. Mode.Parameter.Presel.IReq
id1, id1&2, all, OFF
Mode.Parameter.Presel.SReq
value, unit, all, OFF
Parameters for DET, MET, SET, KFT, STAT, DOS, DOC, MEAS: Automatic in-
quiry after the start of the determination. From such an inquiry, the determina-
tion continues if the requested entry/entries is/are made, e.g.
&SmplData.OFFSilo.Id1 (see 3.2.2.122) or with &M $G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.27. Mode.Parameter.Presel.LimSmplSize.Status ON,
OFF
Mode.Parameter.Presel.LimSmplSize.LoLim
0.0...999 999
Mode.Parameter.Presel.LimSmplSize.UpLim 0.0...999 999
Limit control for the sample size.
3.2.2.28. Mode.Parameter.Presel.ActPuls ON, OFF
for SET, KFT: first, all, cond., OFF
Output of a pulse on the I/O line "Activate", see page 224.