5.1 Setting up and connecting the instruments
799 GPT Titrino
211
inputted from the balance. For this, the address of the identifications
and method, resp. must be preselected on the input unit.
Balance
Method
Id1 Id2 Id3
Sartorius
METH or 27
ID.1 or 26
ID.2 or 24
C-20 or 23
Mettler (AT)
D (Mthd)
C (ID#1) B
(ID#2)
A (c20)
If balance and printer are connected at the same Titrino COM you
need the 6.2125.010 and 6.2125.030 Adapters.
If the balance works only with 7 bit and the printer with 8 bit and if
hey are at the same Titrino COM, the balance has to be set to
"space parity" and Titrino/printer to 8 bit, "no parity".
5.1.5 Connection of a Sample Changer
With 6.2141.030 cable (instead of 6.2141.020), two Titrinos can be
connected to the 730 or 760 Sample Changer at the same time.
• The "Remote" socket allows not only connection of a sample
changer but also additional control functions. Pin assignment of
the "Remote" socket and control possibilities, see page 223.
• If a calibration has to be performed with the sample changer, the
calibration parameter "sample changer:" must be set to "ON".
• In connections with the sample changer, "auto start" should be
set to "OFF" in the
by the Sample Changer.
730/760
6.2141.020 Cable
C
B
A
D