Metrohm 799 GPT Titrino User Manual
Page 176
3.2. Remote control commands
799 GPT Titrino
172
3. For writing onto the display, the standard character set will be used.
4. Unlock the display, see 3.2.2.163
5. Delete the whole display (.DelAll).
6. Send a value to nod &Config.Aux.ResDisplay (see 3.2.2.105) to refresh
the display.
$Q sends the contents of the corresponding display line.
3.2.2.148. Info.ActualInfo.Comport.Number read
only
$Q sends the comport number of the Titrino where the PC is connected.
3.2.2.149. Info.Assembly.CycleTime read
only
Info.Assembly.ExV read
only
Inquiries regarding basic variables of the assembly: Cycle time in s, volume of
the active Exchange Unit in mL.
3.2.2.150. Info.Assembly.Bur.Select internal
D0, external D1, external D2
Selection of buret for assembly functions.
3.2.2.151. Info.Assembly.Bur.Empty
$G, $S, $H, $C
Starts the function "empty". Only possible with the buret type "Dosino". The
parameters are under the function &Config.DosPrep, see 3.2.2.118 up to
3.2.2.120.
3.2.2.152. Info.Assembly.Bur.Prep
$G, $S, $H, $C
Starts the function "preparation". The parameters are under the function
&Config.DosPrep, see 3.2.2.110 up to 3.2.2.120.
3.2.2.153. Assembly.Bur.Rates.Forward.Selected digital, analog
Assembly.Bur.Rates.Forward.Digital 0...150,
max.
Assembly.Bur.Rates.Reverse.Selected digital, analog
Assembly.Bur.Rates.Reverse.Digital 0...150,
max.
Expel and aspirating rate.
Digital or analog control. With digital control, the inputted value applies (in
mL/min). "max." means maximum possible rate with the Exchange Unit in
current use.
Analog means rate control with the analog potentiometer on Titrino.
3.2.2.154. Assembly.Bur.Fill
$G, $H, $C
$G starts the 'FILL' mode of the buret function.
3.2.2.155. Assembly.Bur.ModeDis
$G, $S, $H, $C
Assembly.Bur.ModeDis.Selected volume, time
Assembly.Bur.ModeDis.V 0.0001...0.1...9999
Assembly.Bur.ModeDis.Time 0.25...1...86400
Assembly.Bur.ModeDis.VStop 0.0001...9999, OFF
Assembly.Bur.ModeDis.AutoFill ON, OFF