8rs 232 test, 2 test of the remote box – Metrohm 799 GPT Titrino User Manual
Page 204
4.2. Diagnosis
799 GPT Titrino
200
• Remove test plug.
• Press
diagnose press key 0...9
7.2 Test of the remote box
The remote box needs to be connected before the Titrino is switched on. Insert 3.496.8550 test plug in the remote
box.
• Press <2>.
reset remote box?
• Press
Test
sequence:
1. In case of an error an error message is displayed (for example
error 490 01HEX).
2. If no error is found, after about 1 s display shows
remote box test o.k.
• Press
extern input/output test
• If this test should not be carried out (see 7.1), quit block 4 pressing
ce.
• Remove test plug.
diagnose press key 0...9
8
RS 232 test
A 3.496.8560 test plug normally used in the repair service is required for this test. However, this plug can also be
purchased by customers under the above number.
For the sake of completeness, the procedure is described here. If a diagnostic test of the RS 232 interface is not re-
quired, continue with point 9.
1
2
3
4
5
6
7
8
9
Fig. 5
Connections in the 3.496.8560 plug
• Press <5>.
RS232 test 1...2