Metrohm 799 GPT Titrino User Manual
Page 174
![background image](/manuals/586575/174/background.png)
3.2. Remote control commands
799 GPT Titrino
170
7
pin 12
7
Pulse for recorder (pin 2)
8
not used
(pin 6)
9
not used
(pin 7)
10
Monitoring (pin
8)
11
Monitoring (pin
13)
12
Monitoring (pin
19)
13
Monitoring (pin
20)
3.2.2.141. Info.ActualInfo.Assembly.CyclNo read
only
$Q sends the current cycle number of the voltage measurement cycle, e.g.
"127". From the cycle number and the cycle time (see 3.2.2.149), a time
frame can be set up.
The cycle number is set to 0 on switching on the instrument, on every start
and for QuickMeas. It is incremented as long as the instrument remains
switched on.
3.2.2.142. Info.ActualInfo.Assembly.Counter.V read
only
Info.ActualInfo.Assembly.Counter.Clear $G
$Q sends the volume. With the function &Info.Assembly.Counter.Clear $G, the
volume counter is set to zero.
3.2.2.143. Info.ActualInfo.Assembly.Meas read
only
$Q sends the current measured value from the assembly.
3.2.2.144. Info.ActualInfo.Titrator.CyclNo read
only
Info.ActualInfo.Titrator.V read
only
Info.ActualInfo.Titrator.Meas read
only
Info.ActualInfo.Titrator.dVdt read
only
Info.ActualInfo.Titrator.dMeasdt read
only
Info.ActualInfo.Titrator.dMeasdV read
only
Info.ActualInfo.Titrator.ERC read
only
Info.ActualInfo.Titrator.T read
only
$Q sends the current values in the following formats:
DET
MET
SET
KFT
STAT,
DOC
DOS MEAS
CAL
CyclNo
127 127 127 127 127 127 127 127
V(ml)
1.2345 1.2345 1.2345 1.2345 1.2345 1.2345 -
-
Meas:
pH
U, Ipol (mV)
Upol (uA)
T (°C)
3.345
-345.6
-12.5
-
3.345
-345.6
-12.5
-
3.6(mV)
-345.6
-12.5
-
-
-345.6
-12.5
-
3.6(mV)
-345.6
-12.5
-
3.6(mV)
-345.6
-12.5
3.345
-345.6
-12.5
25.0
3.345
-
-
-
dVdt
(ul/s)
-
-
2.5142 2.5142 2.5142 -
-
-
dMeasdt
pH,U,Ipol mV/s
Upol (uA/s)
T (°C/s)
0.7957
0.7957
-
0.7957
0.7957
-
0.7957
0.7957
-
0.7957
0.7957
-
0.7957
0.7957
-
-
-
-
0.7957
0.7957
0.7957
0.7957
-
-
dMeasdV
(mV/ul)
- - 10.6326 10.6326 10.6326 - - -