2.6 Parameters, key
799 GPT Titrino
63
max.rate max. ml/min
**titr.
min.rate 5.0 µl/min
**titr.
direction: auto
start V: OFF
start V 0.0 ml
factor 0
dos.rate max. ml/min
**titr.
pause 0 s
time interval 10 s
Maximum rate (0.01...150 mL/min, max.)
This parameter primarily determines the addition rate
outside the control range, see also page 68.
The maximum rate depends on the Exchange Unit:
Exchange Unit
max.
5 mL
15 mL/min
10 mL
30 mL/min
20 mL
60 mL/min
50 mL
150 mL/min
Minimum rate (0.01...9999 µL/min)
This parameter determines the rate within the control
range.
Direction is set automatically (+, -, auto)
auto: The direction is set automatically by the Titrino
[U
init
-U
end
]).
+:
Direction of higher pH, higher voltage (more
"positive"), higher current.
- :
Direction of lower pH, lower voltage, lower
currents.
The direction is used only when "init" has been selected
as start value.
Type of start volume (OFF, abs., rel.)
"OFF":
start volume switched off
"abs":
absolute start volume in mL
"rel.":
relative start volume to sample
If "abs." is set:
Absolute start volume (0...999.99 mL)
If "rel." is set:
Factor for relative start volume (0...±999 999)
Calculated as: Start V (in mL) = factor∗sample size
Dosing rate for start volume (0.01...150 mL/min, max.)
The maximum rate depends on the Exchange Unit:
Exchange Unit
max.
5 mL
15 mL/min
10 mL
30 mL/min
20 mL
60 mL/min
50 mL
150 mL/min
Waiting time (0...999 999 s)
Waiting time, e.g. for stabilization of the electrode after
start or reaction time after dosing of a start volume. The
waiting time can be aborted with
Time interval (1...999 999 s)
Time interval for the entry of the measured values in the
measuring point list. The measuring point list can
contain max. 500 points.