3.2 Remote control commands
799 GPT Titrino
129
*Parameter
Tree part "Parameters for TIP"
├
.
Sequence
Sequence
│ ├
.1
Step 1
│ │ ├
.Select
Step selection
method,pause,L4 output,L6 output,
│ │
info,prep,stirrer,OFF 3.2.2.74.
│ │ ├
.Method
Method from mem.or card
special
3.2.2.75.
│ │ ├
.Pause Waiting
time
0...999 999, INF
ditto
│ │ ├
.L4Output Line
L4
active,inactive,pulse,OFF ditto
│ │ ├
.L6Output Line
L6
active,inactive,pulse,OFF ditto
│ │ ├
.Info
Display information
up to 16 ASCII char.
ditto
│ │ ├
.Prep
Preparation of titrating buret
internal D0, external D1,
│ │ │
external D2
ditto
│ │ ├
.Stirrer Stirrer
ON, OFF
ditto
│ ├
:
up to 30 steps
│ │
├
.
Statistics
Statistics
│ ├
.Status
Status of statistics calculation
ON, OFF 3.2.2.24.
│ ├
.MeanN
No. of individual determinations
2...20 ditto
│ ├
.ResTab Result
table
│ │ ├
.Select
original,delete n,delete all
ditto
│ │ ├
.DelN
Deletion of individual results
1...20 ditto
│ │
├
.
Presel
Preselections
│ ├
.IReq
Request of Id's after start
id1, id1&2, all, OFF 3.2.2.26.
│ ├
.SReq
Request of sample size after start
value, unit, all, OFF ditto
│ ├
.LimSmplSize
Limits for sample size
3.2.2.27
│ │ ├
.Status
Status of limit control
ON, OFF ditto
│ │ ├
.LoLim Lower
limit
0.0...999 999
ditto
│ │ ├
.UpLim Upper
limit
0.0...999 999 ditto
│ ├
.MeasMode
Measuring mode for man.meas. pH,U,Ipol,Upol,T,OFF 3.2.2.76.
│ ├
.MeasInput Measuring
input
1, 2, diff.
ditto
│ ├
.Ipol
Polarization
current
0...1...±127 ditto
│ ├
.Upol
Polarization voltage
0...400...±1270 ditto
│ ├
.PolElectrTest
Test for polarized electrodes
ON, OFF ditto
│ ├
.Temp
Titration temperature
-170.0...25.0...500.0 ditto