beautypg.com

Metrohm 799 GPT Titrino User Manual

Page 48

background image

2.6. Parameters, key

799 GPT Titrino

44


min.rate 25.0 µl/min
**titr.



Minimum dosing rate (0.01...9999 µL/min)

This parameter determines the addition rate in the

range of the control point.

Rule of thumb for "min.rate" in µL/min = (expected rate

of the reaction in µL/min)/10.


>titration parameters

start V: OFF





start V 0.0 ml


factor 0


dos.rate max. ml/min

**titr.









pause 0 s

**titr.




start time 0 s



start pH OFF














Titration parameters

Type of start volume (OFF, abs., rel.)

"OFF":

start volume switched off

"abs.":

absolute start volume in mL

"rel.":

relative start volume to sample size.

If "abs." is set:

Absolute start volume (0...999.99 mL)

If "rel." is set:

Factor for relative start volume (0...±999 999).

Calculated as: start V (in mL) = factor∗sample size

Dosing rate for start volume

(0.01...150 mL/min, max.)

sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Waiting time after start volume (0...999 999 s)

Waiting time, e.g. for equilibration of the electrode after

start or reaction time after dosing of a start volume. The

waiting time can be aborted with .

Start time for data acquisition (0...999 999 s)

Measured values will be acquired only after the start

time has elapsed.

Start measured value for data acquisition (input range

depends on the measured quantity:

pH: 0...±20.00,

OFF

U, Ipol: 0...±2000 mV, OFF

Upol:

0...±200.0 µA, OFF)

sets "OFF".

Measured values will be acquired only after start of

measured value has been reached.