beautypg.com

Metrohm 799 GPT Titrino User Manual

Page 33

background image

2.6 Parameters, key

799 GPT Titrino

29

req.smpl size: OFF

cond.



limit smpl size: OFF

cond.






low lim. 0.0

cond.

up lim. 999999

cond.


activate pulse: OFF

cond.

Request of sample size after start of titration (value, unit,

all, OFF)

"all" the value and the unit will be requested.

Limiting value check for sample size (ON, OFF)

With "on" the error message "sample size out." appears

if the entry is outside the set limits. The limiting values

are shown in the display window.

The absolute value of the limit is checked during sam-

ple size input and during the calculation of the results.

If "on" has been set:

Lower limit for sample size (0.0...999 999)

Upper limit for sample size (0.0...999 999)

Pulse output on I/O line 6 (L6, pin 1) of the remote

socket (first, all, cond., OFF)

see page 224.